Test 1: Mechanical
Inspection
Each load cell should be inspected for physical damage.
A thorough inspection of the entire length of any cables should be done.
Nicked or abraded cables can short out a load cell. Inspect for distortion
or cracks on all metal surfaces. Flexure surfaces, if any, should be parallel
to each other and be perpendicular to both end surfaces. If the load cell
has been physically deformed (bent, stretched, or compressed from original
configuration) it is non-repairable. Consult Aerocon for load cell application
information.
Test 2: Zero Balance
(electrical output with no load)
Changes in the zero balance are usually caused by residual
stress in the sensing area. Residual stresses result from overloading the
cell, and /or from millions of cycles.
With a millivoltmeter, measure the load cell's output
under a 'no load' condition, (no weight on the load cell). The output of
a trimmed cell should typically be within +/- .1% of rated output. Always
consult published specifications to obtain the correct values.
EXAMPLE: If the typical load cell output =2 mV/V
and the typical power supply output =10V, then the full scale output would
be =20 mV. Typical zero balance =+/-.1% of rated output. The zero reading
should then be within +/- .02 mV.
If the output is greater than the zero balance tolerance,
the cell is damaged but may still be useable. When the zero reading exceeds
the zero balance tolerance but is less than 50% of the full scale output,
the load cell may be electronically 're-zeroed' if the electronic readout
with the cell allows it; however, there is no guarantee that the load cell
will remain linear. If a load cell has more than 50% full scale zero balance
shift, it may require 're-gauging'.
Test 3: Bridge
Resistance
Out of tolerance bridge resistance values are usually
caused by the failure of a compensating element. Another cause is a broken
or burned bridge wire. These failures commonly result from electrical transients
or lightening strikes.
TO TEST FOR BRIDGE RESISTANCE: Measure the resistance
across each pair of input and output leads. The input and output
resistance is typically 350 +/- 3.5 ohms, however, there are many variations,
depending on each load cell's published specifications. Consult the published
specifications of the subject load cell for proper values and tolerances.
If these readings are 'out of spec', the load cell requires repair.
Test 4: Resistance
to Ground
Current flow from the load cell's electronics to the
load cell body is usually caused by water contamination within the load
cell or its cable.
TO TEST FOR LEAKAGE: Connect all of the input, output,
sense and ground leads together and measure the resistance between the
load cell body and the leads with a megohmmeter. Make sure good contact
is made with the metal load cell sensor. The reading should be at least
5000 Megohms. If the load cell fails this test, retest without the ground
wire. If the load cell passes this test, a problem may exist with the load
cell cable. If the test fails, the load cell requires repair.